Analytical Method Development and Validation RP-HPLC method for Estimation of Tadalafil in Bulk and Oral Jelly Formulation
Mohammad Aman Hakimoddin*, Aejaz Ahmed, Khan Javed, Manyar Abuzaid Muzaffar
Department of Pharmaceutical Chemistry, JIIU’S Ali-Allana College of Pharmacy Akkalkuwa, Dist: Nandurbar.
*Corresponding Author E-mail: anamshaikh2805@gmail.com
ABSTRACT:
1. INTRODUCTION:
In order to identify, characterize, and quantify pharmaceutical compounds and their formulations, analytical chemistry is essential1. Because of its great resolution, sensitivity, and reproducibility, high-performance liquid chromatography (HPLC) is a commonly used analytical method2. In the pharmaceutical sector, reverse-phase HPLC (RP-HPLC) has become the preferred technique for routine drug substance analysis3–5.
A strong and specific phosphodiesterase type 5 (PDE5) inhibitor, tadalafil is prescribed to treat pulmonary arterial hypertension (PAH), benign prostatic hyperplasia
(BPH), and erectile dysfunction (ED)6–9. It works by preventing cyclic guanosine monophosphate (cGMP) from degrading, which promotes smooth muscle relaxation and vasodilation10. Tadalafil is appropriate for once-daily dosage regimens because of its sustained pharmacologic effects and lengthy half-life (~17.5 hours)11–12.
Oral jelly formulations have been more well-liked recently as a cutting-edge drug delivery method. For elderly, paediatric, and dysphagic patients who have trouble swallowing traditional solid dosage forms, they are particularly helpful13–15. These formulations provide improved palatability, quicker onset of action, and increased patient compliance. Nevertheless, there aren't many analytical techniques in the literature that are specifically designed to evaluate the stability and content homogeneity of such semi-solid systems.
Prior research has shown that RP-HPLC may be used to estimate the amount of tadalafil in biological matrices, pills, and bulk medications16–19. However, due to the presence of gelling agents, sweeteners, and preservatives, oral jelly formulations provide unique matrix problems, and only a small number of methodologies have been described for their examination20–22.
Thus, the goal of the current study is to create and validate a straightforward, reliable, and specific RP-HPLC method for measuring tadalafil in oral jelly and bulk medication formulations. In order to ensure its application in routine quality control analysis and regulatory compliance, the developed approach has been validated in accordance with ICH Q2 (R1) requirements23–24.
2. MATERIALS AND METHODS:
2.1 Drug Profile:
The PDE5 inhibitor tadalafil was used to treat pulmonary arterial hypertension, benign prostatic hyperplasia, and erectile dysfunction. Oral administration was used, and dosages varied depending on the situation. The medicine worked by inhibiting PDE5, which raised cGMP levels to encourage vasodilation and smooth muscle relaxation. Rapid absorption, CYP3A4 metabolism, and a half-life of approximately 17.5 hours were among its pharmacokinetic characteristics. Hypersensitivity and co-administration of nitrate were among the contraindications. Headache, dyspepsia, and flushing were common side effects; priapism and vision or hearing losses were uncommon but dangerous side effects25.
2.2 Chemicals and Reagents:
The tadalafil reference standard was obtained from a certified pharmaceutical supplier with stated purity above 99% 24. Methanol (HPLC grade), orthophosphoric acid, and double-distilled water were used for mobile phase preparation. All chemicals and reagents used in this study were of analytical or HPLC grade and were procured from Merck or equivalent standard sources. Jelly formulation of tadalafil was purchased as Apcalis sx of Ajanta Pharma25.
2.3 Instrumentation and Chromatographic Conditions:
An HPLC device with a quaternary gradient pump, a manual injector with a 20 µL fixed loop, and a UV-visible detector made up the chromatographic system. Software for chromatography that was compatible was used to acquire data. 26
· Column: C18 reverse phase (250 mm × 4.6 mm i.e, 5 µm particle size)
· Mobile Phase: Methanol: Water in a ratio of 60:40 (v/v), adjusted to pH 3.5 using orthophosphoric acid
· Flow Rate: 1.0 mL/min
· Detection Wavelength: 286 nm
· Injection Volume: 20 µL
· Column Temperature: Ambient (25 ± 2°C)
· Run Time: 10 minutes
The mobile phase was degassed and filtered through a 0.45 µm membrane filter prior to use27.
2.4 Preparation of Standard Solution:
A standard stock solution of tadalafil was prepared by dissolving an accurately weighed quantity of the drug in methanol to obtain a concentration of 100 µg/mL. From this stock, working standard solutions were prepared by serial dilution to achieve concentrations of 5, 10, 15, 20, and 25 µg/mL28.
2.5 Sample Preparation (Oral Jelly Formulation):
Accurately weighed quantity of tadalafil oral jelly equivalent to 20 mg of tadalafil was transferred to a 100 mL volumetric flask, dissolved in methanol, sonicated for 15 minutes, and filtered. The filtrate was further diluted to obtain a final concentration of 10 µg/mL29.
2.6 Method Validation:
The following parameters were used to validate the approach in accordance with ICH Q2(R1) criteria:
· Linearity: assessed for a range of concentrations (5–25 µg/mL; n = 6). Peak area versus concentration was used to plot the calibration curve30.
· Accuracy: Ascertained using the usual addition method at three levels: 80%, 100%, and 120%31.
· Precision: Results are given as a percentage of RSD and are evaluated by intra- and inter-day variance.
· Specificity: Verified by observing interference from jelly excipients during the tadalafil retention period.
· The formulas LOD = 3.3(σ/S) and LOQ = 10(σ/S) were used to calculate LOD and LOQ, where S is the slope and σ is the response's standard deviation32.
· Robustness: Assessed by slight variations in wavelength (±2 nm), pH (±0.2), and flow rate (±0.1 mL/min).
· System Suitability: Parameters like retention time, theoretical plates, tailing factor, and capacity factor were checked33.
3. RESULTS AND DISCUSSION:
3.1 System Suitability:
Prior to sample analysis, system suitability tests were conducted to guarantee the chromatographic system's reproducibility34–35. Tadalafil was found to have an average retention time of 5.2 ± 0.1 minutes36. Good column efficiency was indicated by the theoretical plate count of >3000, and adequate peak symmetry was demonstrated by the tailing factor of <1.537–38. The ICH guidelines' acceptance criteria were satisfied by these characteristics39.
|
Sr. No. |
Column |
Mobile Phase |
Flow Rate (mL/min) |
Observation |
|
1 |
C18, 250×4.6 mm, 5 µm |
MeOH 90% + 0.1% Acetic Acid 10% |
0.7 |
Early elution, broad peak, very low plate count |
|
2 |
C18, 250×4.6 mm, 5 µm |
MeOH 80% + 0.1% Acetic Acid 20% |
0.7 |
Two peaks, better resolution, high plate number |
|
3 |
C18, 250×4.6 mm, 5 µm |
MeOH 70% + 0.1% Acetic Acid 30% |
0.7 |
High response, sharp peak, good efficiency |
|
4 |
C18, 250×4.6 mm, 5 µm |
MeOH 60% + 0.1% Acetic Acid 40% |
0.7 |
Further improved area, but peak width slightly broader |
|
5 |
C18, 250×4.6 mm, 5 µm |
MeOH 60% + 0.1% Acetic Acid 40% |
0.8 |
Good peak shape, reduced retention time slightly |
|
6 |
C18, 250×4.6 mm, 5 µm |
MeOH 60% + 0.1% Acetic Acid 40% |
0.8 |
sharp peak, good plates, symmetry |
Fig 1.2: Linearity
Table 1: System Suitability Parameters for Tadalafil
|
Parameter |
Observed Value |
Acceptance Criteria |
|
Retention Time (Rt) |
5.2 min |
Consistent (± 2%) |
|
Theoretical Plates (N) |
> 3000 |
> 2000 |
|
Tailing Factor (T) |
< 1.5 |
≤ 2.0 |
|
Capacity Factor (k') |
3.1 |
> 2 |
3.2 Linearity:
The technique demonstrated a linear response within the 5–25 µg/mL concentration range40. Excellent linearity was indicated by the calibration curve's regression equation of y = 98745x + 2356 and correlation coefficient (r2) of 0.99941–42.
Table 2: Linearity Data for Tadalafil
|
Concentration (µg/mL) |
Mean Peak Area |
%RSD |
|
5 |
492,200 |
0.86 |
|
10 |
982,400 |
0.74 |
|
15 |
1,478,600 |
0.69 |
|
20 |
1,965,000 |
0.52 |
|
25 |
2,459,500 |
0.61 |
3.3 Accuracy:
Accuracy was assessed through recovery studies at three concentration levels: 80%, 100%, and 120%. The mean recovery ranged from 99.2% to 101.5% 43–44.
Fig 1.3 Chromatogram of Accuracy 80% - 01
Fig 1.5 Chromatogram of Accuracy 80% - 02
Fig 1.7 Chromatogram of Accuracy 100% - 01
Fig 1.4 Chromatogram of Accuracy 100% - 02
Fig 1.6 Chromatogram of Accuracy 120% - 01
Fig 1.8 Chromatogram of Accuracy 120% - 02
Table 3: Recovery Studies (Accuracy)
|
Level (%) |
Amount Added (µg/mL) |
Amount Recovered (µg/mL) |
% Recovery |
% RSD |
|
80 |
8.0 |
7.96 |
99.5 |
0.85 |
|
100 |
10.0 |
10.13 |
101.3 |
0.72 |
|
120 |
12.0 |
11.92 |
99.2 |
0.68 |
3.4 Precision:
Precision was evaluated at three different concentrations for both intra-day and inter-day variations. Results confirmed acceptable repeatability and reproducibility45–46.
Fig 1.9 Chromatogram of Precision (Intraday) 10 μg - 01
Fig 1.11 Chromatogram of Precision (Intraday)10 μg – 02
Fig 1.13 Chromatogram of Precision (Intraday)15 μg – 01
Fig 1.10 Chromatogram of Precision (Intraday) 15 μg – 02
Fig 1.12 Chromatogram of Precision (Intraday) 20 μg – 01
Fig 1.14 Chromatogram of Precision (Intraday) 20 μg – 02
Table 4: Precision Data (n = 3)
|
Concentration (µg/mL) |
Intra-day %RSD |
Inter-day %RSD |
|
10 |
0.72 |
1.04 |
|
15 |
0.85 |
1.21 |
|
20 |
0.65 |
0.97 |
3.5 LOD and LOQ:
LOD and LOQ values were determined using standard deviation of the response and slope method47–48.
Table 5: Sensitivity Parameters
|
Parameter |
Value (µg/mL) |
|
LOD |
0.45 |
|
LOQ |
1.35 |
3.6 Robustness:
The robustness of the method was verified by applying small changes in chromatographic conditions. No significant change was observed in retention time, peak area, or resolution49–50.
Table 6: Robustness Study Results
|
Parameter Changed |
Modified Value |
Rt (min) |
% RSD |
|
Flow Rate |
0.9 mL/min |
5.34 |
0.89 |
|
1.1 mL/min |
5.01 |
0.92 |
|
|
Wavelength |
284 nm |
5.20 |
0.85 |
|
288 nm |
5.18 |
0.88 |
|
|
Mobile Phase (MeOH:H₂O) |
58:42 and 62:38 (v/v) |
5.22 |
0.76 |
Fig 1.15 Chromatogram of Assay of Marketed Formulation – 01
Fig 1.16 Chromatogram of Assay of Marketed Formulation – 02
Table 7: Chromatogram of Assay of Marketed Formulation
|
Trial |
Ret Time (min) |
Area (mAU*s) |
Height (mAU) |
Symmetry |
Width (min) |
Plates |
|
Trial 01 |
3.886 |
772.13538 |
78.00574 |
0.46 |
0.1427 |
4104 |
|
Trial 02 |
3.894 |
764.77155 |
77.53973 |
0.46 |
0.1438 |
4070 |
|
Mean |
— |
768.45347 |
— |
— |
— |
— |
|
%RSD |
— |
0.48% |
— |
— |
— |
— |
4. CONCLUSION:
The quantification of tadalafil in bulk and oral jelly formulations was accomplished by the successful development and validation of a novel, straightforward, and reliable reverse-phase high-performance liquid chromatography (RP-HPLC) method. The technique offers high sensitivity and selectivity quantification that is quick and accurate. It exhibits outstanding linearity, precision, accuracy, specificity, and resilience while adhering to ICH Q2(R1) requirements.
Under the ideal chromatographic circumstances, tadalafil demonstrated a consistent retention time and acceptable system compatibility properties. The method's robustness and ruggedness were confirmed when it was shown to be insensitive to small changes in analytical parameters. Its usefulness for practical pharmaceutical analysis is confirmed by its effective application to an oral jelly matrix free from excipient influence.
The suggested RP-HPLC method can be regularly employed in the quality control and stability investigations of tadalafil formulations due to its efficacy and regulatory compliance, particularly for novel dosage forms as oral jellies. After the required adjustments, it also shows potential for additional adaption in pharmacokinetic or bioanalytical investigations.
5. ACKNOWLEDGEMENTS:
Thankful for encouragement and support provided President Moulana Huzaifa Vastanvi and Principal Ali–Allana College of pharmacy, Akkalkuwa Dist- Nandurbar, Maharashtra India. We also wish to sincerely thank the Swapnroop Drugs and Pharmaceuticals Ltd for providing API.
6. CONFLICT OF INTEREST:
The authors have declared no conflict of interest.
7. REFERENCES
1. ICH Q2 (R1). Validation of Analytical Procedures: Text and Methodology. International Conference on Harmonisation; 2005.
2. Skoog DA, Holler EJ, Crouch SR. Principles of Instrumental Analysis. 6th ed. Thomson Brooks/Cole; 2007.
3. Chatwal GR, Anand SK. Instrumental Methods of Chemical Analysis. 5th ed. Himalaya Publishing House; 2004.
4. Khopkar SM. Basic Concepts of Analytical Chemistry. 3rd ed. New Age International; 2008.
5. Christian GD. Analytical Chemistry. 6th ed. John Wiley and Sons; 2007.
6. Green MJ. A practical guide to analytical method validation. Chem News Features. 1996; 2(4): 305-10.
7. Sharma BK. Instrumental Methods of Chemical Analysis. Goel Publishing House; 2000.
8. Ludwig H, Huber L. Validation of Analytical Methods: Review and Strategy. Agilent Technologies; 2001.
9. Chandran S, Singh R. Comparison of various international guidelines for analytical method validation. Pharm Rev. 2007;5(4):12-14.
10. Snyder LR, Kirkland JJ, Glajch JL. Practical HPLC Method Development. 2nd ed. Wiley; 1997.
11. Abidi SL. HPLC of phosphatidic acids and related polar lipids. J Chromatogr. 1991; 587: 193-203.
12. Hearn MTW. Ion-pair chromatography on normal and reversed-phase systems. Adv Chromatogr. 1980; 18: 59–100.
13. Sonawane AA et al. Area under Curve UV Spectrophotometric Method for Determination of Tadalafil in Bulk. Bull Env Pharmacol Life Sci. 2020; 9(9): 12–6.
14. Kumar YS et al. Development and Validation of RP-HPLC Method for Estimation of Tadalafil in Bulk and Tablet Dosage Form. J Chem Pharm Res. 2012;4(3):1783–87.
15. Patel JK, Patel NK. Stability-Indicating RP-HPLC Method for Determination of Ambrisentan and Tadalafil. Sci Pharm. 2014;82(4):749–63.
16. Salgar J et al. Development and validation of RP-HPLC method for estimation of Tadalafil in tablet dosage form. Int J Pharm Pharm Sci. 2016; 8(8): 356–8.
17. Sathyaraj A et al. Development and validation of RP-HPLC method for estimation of Tadalafil in pharmaceutical formulations. Int J Pharm Adv Res. 2019; 8(2).
18. Unnisa A et al. RP-HPLC-PDA method for the analysis of Tadalafil in bulk and formulations. Int J Pharm Pharm Sci. 2014; 6(6): 308–12.
19. Amin G et al. Dual wavelength spectrophotometric method for estimation of Tadalafil and Depoxitine HCl. Int J Pharm Res Bio-Sci. 2012;1(2):247–55.
20. Khan ZG et al. Spectrophotometric methods for estimation of Tadalafil. Int J Pharm Pharm Sci. 2014;6(2):662–5.
21. Reddy BP et al. Stability-indicating RP-HPLC method for Tadalafil. Int J Pharm Sci Res. 2013;4(6):2331–6.
22. Raul SK et al. RP-HPLC method for estimation of Tadalafil. Res J Pharm Technol. 2012;5(12):1573–6.
23. Goldstein I et al. Oral Tadalafil for Erectile Dysfunction. N Engl J Med. 2003;349(20):1907–15.
24. Fraihat S. Spectrophotometric Methods for Determination of Tadalafil. Int J Pharm Pharm Sci. 2014;6(7):443–5.
25. Yunoos M, Sankar DG. UV Spectrophotometric Determination of Tadalafil. E-J Chem. 2010;7(3):833–6.
26. Yada D et al. RP-HPLC method development for Tadalafil. Int J ChemTech Res. 2010;2(1):329–33.
27. Bojanapu A et al. Validation of RP-HPLC for Tadalafil estimation. Drug Res. 2010; 65:82–5.
28. Samala A et al. RP-HPLC estimation of Tadalafil in pharmaceutical dosage form. Res Pharm Biotechnol. 2010;2(1):1–6.
29. Rabbaa-Khabbaz L et al. Determination of Tadalafil in human serum by HPLC. J Appl Res. 2006;6(1):62–7.
30. Al AR et al. RP-HPLC method for Esomeprazole and Tadalafil estimation. Int J Pharm Pharm Sci. 2013;5(3):638–41.
31. Ahmed R et al. Spectrophotometric determination of Tadalafil in wastewater. J Anal Chem. 2013;68(3):257–62.
32. Chenthilnathan A, Shanmugasundaram P. RP-HPLC estimation of Tadalafil and Depoxitine. Int J Pharm Biol Sci. 2014;4(2):72–82.
33. Khan SA et al. Analytical method validation for PDE5 inhibitors in herbal sachets. J Chrom Sci. 2023;61(4):313–9.
34. Ali HSM et al. Sono-assisted nanoemulsion oral jellies of Tadalafil. Drug Dev Ind Pharm. 2022;48(2):167–77.
35. Prasanna Reddy B et al. Stability-indicating RP-HPLC for Tadalafil. Pharm Chem J. 2013;47(6):441–7.
36. Sadhana J et al. UV-spectrophotometric and HPLC methods for Ambrisentan and Tadalafil. Indian J Pharm Sci. 2014;76(3):254–60.
37. Salunkhe S et al. RP-HPLC for Tadalafil and combination drugs. Int J Pharm Pharm Sci. 2021;67(2):126–30.
38. Dhaneshwar SR et al. Simultaneous HPLC of Tadalafil and Depoxitine. Int J Chem Res. 2012;4(3):189–94.
39. Jensen W. Melting Points and Organic Compounds. J Chem Educ. 2009;86.
40. Indian Pharmacopoeia, 2022. Organoleptic Properties. Govt. of India.
41. British Pharmacopoeia, 2023. Appendix XXII A.
42. Drugs.com. Tadalafil (Professional Monograph).
43. Eli Lilly Canada Inc. Cialis® Product Monograph.
44. Kloner RA, Zusman RM. Cardiovascular effects of PDE5 inhibitors. Am J Cardiol. 2011;107(8):1202–8.
45. US FDA. Technical Review Guide: Chromatographic Method Validation. 1993.
46. Synder LR et al. Practical HPLC Development. Wiley; 1997.
47. Christian G. Analytical Chemistry. John Wiley; 2007.
48. Huber L. Validation of Analytical Methods. Lab Compliance; 2001.
49. Riley CM, Rosanske TW. Development and Validation of Analytical Methods. Elsevier; 1996.
50. Somenath M. Sample Prep Techniques in Analytical Chemistry. Wiley-Interscience; 2003.
|
Received on 25.06.2025 Revised on 21.02.2026 Accepted on 24.07.2026 Published on 12.09.2026 Available online from September 19, 2026 Asian J. Research Chem.2026; 19(5):405-411. DOI: 10.52711/0974-4150.2026.00060 ©A and V Publications All Right Reserved
|
|
|
This work is licensed under a Creative Commons Attribution-Non Commercial-Share Alike 4.0 International License. Creative Commons License. |
|